Section6: SCPI command referenceModel 2461 Interactive SourceMeter® Instrument Reference Manual6-2382461-901-01 A/November 2015:TRIGger:TIMer:DELayThis command sets and reads the timer delay.TypeAffected byWhere savedDefault valueCommand and queryRecall settingsInstrument resetPower cycleSave settings10e-6 (10μs)Usage:TRIGger:TIMer:DELay:TRIGger:TIMer:DELay?Trigger timer number (1 to 4)Delay interval (8e-6 s to 100,000 s)DetailsEach time the timer is triggered, it uses this delay period.Reading this command returns the delay interval that will be used the next time the timer is triggered.ExampleTRIG:TIM2:DEL50E-6Set trigger timer 2 to delay for 50 μs.Also seeNone:TRIGger:TIMer:STARt:FRACtionalThis command configures an alarm or a time in the future when the timer will start.TypeAffected byWhere savedDefault valueCommand and queryRecall settingsInstrument resetPower cycleSave settings0Usage:TRIGger:TIMer:STARt:FRACtional :TRIGger:TIMer:STARt:FRACtional?Trigger timer number (1 to 4)The time in fractional seconds (0 to < 1 s)