Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 8: TSP command reference2461-901-01 A/November 2015 8-305To use the timer as a simple delay or pulse generator with digital I/O lines, make sure the timer starttime in seconds and fractional seconds is configured for a time in the past. To use the timer as analarm, configure the timer start time in seconds and fractional seconds for the desired alarm time.Exampletrigger.timer[3].enable = trigger.ON Enable the trigger timer for timer 3.Also seeNonetrigger.timer[N].reset()This function resets trigger timer settings to their default values.Type TSP-Link accessible Affected by Where saved Default valueFunction YesUsagetrigger.timer[N].reset()N Trigger timer number (1 to 4)DetailsThe trigger.timer[N].reset() function resets the following attributes to their default values:• trigger.timer[N].count• trigger.timer[N].delay• trigger.timer[N].delaylist• trigger.timer[N].enable• trigger.timer[N].start.generate• trigger.timer[N].start.fractionalseconds• trigger.timer[N].start.seconds• trigger.timer[N].stimulusIt also clears trigger.timer[N].overrun.Exampletrigger.timer[1].reset() Resets the attributes associated with timer 1 totheir default values.Also seetrigger.timer[N].count (on page 8-301)trigger.timer[N].delay (on page 8-303)trigger.timer[N].delaylist (on page 8-303)trigger.timer[N].enable (on page 8-304)trigger.timer[N].start.fractionalseconds (on page 8-306)trigger.timer[N].start.generate (on page 8-306)trigger.timer[N].start.overrun (on page 8-307)trigger.timer[N].start.seconds (on page 8-308)trigger.timer[N].start.stimulus (on page 8-308)