Section 6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument Reference Manual6-216 2461-901-01 A/November 2015:TRIGger:LAN:IN:CLEarThis command clears the event detector for a LAN trigger.Type Affected by Where saved Default valueCommand only Not applicable Not applicable Not applicableUsage:TRIGger:LAN:IN:CLEar The LAN event number (1 to 8) to clearDetailsThe trigger event detector enters the detected state when an event is detected. This function clears atrigger event detector and discards the previous of the trigger packet.This function clears all overruns associated with this LAN trigger.Example:TRIG:LAN5:IN:CLE Clears the event detector with LAN packet 5.Also see:TRIGger:LAN:IN:OVERrun? (on page 6-217):TRIGger:LAN:IN:EDGEThis command sets the trigger operation and detection mode of the specified LAN event.Type Affected by Where saved Default valueCommand and query Recall settingsInstrument resetPower cycleSave settings EITHUsage:TRIGger:LAN:IN:EDGE :TRIGger:LAN:IN:EDGE? The LAN event number (1 to 8) The trigger mode; see the Details for more informationDetailsThis command controls how the trigger event detector and the output trigger generator operate on thegiven trigger. These settings are intended to provide behavior similar to the digital I/O triggers.LAN trigger mode valuesMode Trigger packets detected as input LAN trigger packet generated foroutput with a…EITHer Rising or falling edge (positive or negative state) negative stateFALLing Falling edge (negative state) negative stateRISing Rising edge (positive state) positive state