Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI command reference2461-901-01 A/November 2015 6-235The trigger model states are:• Idle: The trigger model is stopped.• Running: The trigger model is running.• Waiting: The trigger model has been in the same wait block for more than 100 ms.• Empty: The trigger model is selected, but no blocks are defined.• Building: Blocks have been added.• Failed: The trigger model is stopped because of an error.• Aborting: The trigger model is stopping because of a user request.• Aborted: The trigger model is stopped because of a user request.Example:TRIG:STAT? An example output if the trigger model isinactive and ended at block 9 is:IDLE;IDLE;9Also seeNone:TRIGger:TIMer:CLEarThis command clears the timer event detector and overrun indicator for the specified trigger timer number.Type Affected by Where saved Default valueCommand only Not applicable Not applicable Not applicableUsage:TRIGger:TIMer:CLEar Trigger timer number (1 to 4)DetailsThis command sets the timer event detector to the undetected state and resets the overrun indicator.Example:TRIG:TIM1:CLEar Clears trigger timer 1.Also see:TRIGger:TIMer:COUNt (on page 6-236):TRIGger:TIMer:STARt:OVERrun? (on page 6-240)