Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 8: TSP command reference2461-901-01 A/November 2015 8-221DetailsThis command selects which set of input and output terminals the instrument uses. You can selectfront panel or rear panel terminals.If the output is turned on when you change from one set of terminals to the other, the output is turnedoff.This command replaces the smu.measure.terminals command, which is deprecated.Examplesmu.terminals = smu.TERMINALS_FRONT Use the front-panel terminals for measurements.Also seeNonestatus.clear()This function clears event registers and the event log.Type TSP-Link accessible Affected by Where saved Default valueFunction YesUsagestatus.clear()DetailsThis command clears the event registers of the Questionable Event and Operation Event Registerset. It does not affect the Questionable Event Enable or Operation Event Enable registers.Examplestatus.clear() Clear the bits in the registersAlso see*CLS (on page B-2)status.conditionThis attribute stores the status byte condition register.Type TSP-Link accessible Affected by Where saved Default valueAttribute (R) Yes Not applicable Not applicable Not applicableUsagestatusByte = status.conditionstatusByte The status byte