Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 8: TSP command reference2461-901-01 A/November 2015 8-243trigger.digin[N].overrunThis attribute returns the event detector overrun status.Type TSP-Link accessible Affected by Where saved Default valueAttribute (R) Yes Digital I/O trigger N clearDigital I/O trigger N resetNot applicable Not applicableUsageoverrun = trigger.digin[N].overrunoverrun Trigger overrun state (true or false)N Digital I/O trigger line (1 to 6)DetailsIf this is true, an event was ignored because the event detector was already in the detected statewhen the event occurred.This is an indication of the state of the event detector built into the line itself. It does not indicate if anoverrun occurred in any other part of the trigger model or in any other detector that is monitoring theevent.Exampleoverrun = trigger.digin[1].overrunprint(overrun)If there is no trigger overrun on digitalinput 1, the output is:falseAlso seedigio.line[N].mode (on page 8-50)digio.line[N].reset() (on page 8-52)Digital I/O port configuration (on page 3-87)trigger.digin[N].clear() (on page 8-241)trigger.digin[N].wait()This function waits for a trigger.Type TSP-Link accessible Affected by Where saved Default valueFunction YesUsagetriggered = trigger.digin[N].wait(timeout)triggered Trigger detected: trueNo triggers detected during the timeout period: falseN Digital I/O trigger line (1 to 6)timeout Timeout in seconds