Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 8: TSP command reference2461-901-01 A/November 2015 8-309Trigger events are described in the table below.Trigger eventsEvent description Event constantNo trigger event trigger.EVENT_NONEFront-panel TRIGGER key press trigger.EVENT_DISPLAYNotify trigger block N (1 to 8) generates a trigger event whenthe trigger model executes ittrigger.EVENT_NOTIFYNA command interface trigger:• Any remote interface: *TRG• GPIB only: GET bus command• VXI-11: VXI-11 command device_triggertrigger.EVENT_COMMANDLine edge (either rising, falling, or either based on theconfiguration of the line) detected on digital input line N(1 to 6)trigger.EVENT_DIGIONLine edge detected on TSP-Link synchronization line N(1 to 3)trigger.EVENT_TSPLINKNAppropriate LXI trigger packet is received on LAN triggerobject N (1 to 8)trigger.EVENT_LANNTrigger event blender N (1 to 2), which combines triggereventstrigger.EVENT_BLENDERNTrigger timer N (1 to 4) expired trigger.EVENT_TIMERNSource limit condition occurs trigger.EVENT_SOURCE_LIMITExampledigio.line[3].mode = digio.MODE_TRIGGER_INtrigger.timer[1].delay = 3e-3trigger.timer[1].start.stimulus = trigger.EVENT_DIGIO3Set digital I/O line 3 to be atrigger input.Set timer 1 to delay for 3 ms.Set timer 1 to start the timerwhen an event is detected ondigital I/O line 3Also seeNonetrigger.timer[N].wait()This function waits for a trigger.Type TSP-Link accessible Affected by Where saved Default valueFunction YesUsagetriggered = trigger.timer[N].wait(timeout)triggered Trigger detection indicationN Trigger timer number (1 to 4)timeout Maximum amount of time in seconds to wait for the trigger