Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 8: TSP command reference2461-901-01 A/November 2015 8-97reset()This function resets commands to their default settings and clears the buffers.Type TSP-Link accessible Affected by Where saved Default valueFunction NoUsagereset()reset(system)system If the node is the master, the entire system is reset: trueOnly the local group is reset: falseDetailsThe reset() command in its simplest form resets the entire TSP-enabled system, including thecontrolling node and all subordinate nodes.If you want to reset a specific instrument, use the node[N].reset() command. Also use thenode[N].reset() command to reset an instrument on a subordinate node.When no value is specified for system, the default value is true.You can only reset the entire system using reset(true) if the node is the master. If the node is notthe master node, executing this command generates an error event.Examplereset(true) If the node is the master node, the entire system isreset; if the node is not the master node, an errorevent is generated.Also seeResets (on page 2-141)localnode.prompts4882This attribute enables and disables the generation of prompts for IEEE Std 488.2 common commands.Type TSP-Link accessible Affected by Where saved Default valueAttribute (RW) Yes Power cycle Not saved localnode.ENABLEUsageprompting = localnode.prompts4882localnode.prompts4882 = promptingprompting IEEE Std 488.2 prompting mode:• Disable prompting: localnode.DISABLE• Enable prompting: localnode.ENABLE