Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 8: TSP command reference2461-901-01 A/November 2015 8-273limit4Pattern The bit pattern that is sent when the measurement passes limit 4; range 1to 15; default is 8allPattern The bit pattern that is sent when all limits have failed; 1 to 15; default is 15startInLine The input line that starts the test; 5 for digital line 5, 6 for digital line 6;default is 5startDelay The delay time before each measurement (167 ns to 10 ks); default is 0 forno delayendDelay The delay time after the measurement (167 ns to 10 ks); default is 0 for nodelaybufferName The name of the reading buffer, which may be a default buffer(defbuffer1 or defbuffer2) or a user-defined buffer; defaults todefbuffer1DetailsThis trigger model template allows you to sort components and place them into up to four bins, basedon the comparison to limits.To set a limit as unused, set the high value for the limit to be less than the low limit.All limit patterns and the all fail pattern are sent on digital I/O lines 1 to 4, where 1 is the leastsignificant bit.After selecting a trigger model template, you can view the trigger model blocks in a graphical formatby pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add ordelete blocks and change trigger model settings from this screen. You can use thetrigger.model.getblocklist() command to view the trigger model blocks in a list format.Also seeNonetrigger.model.setblock() — trigger.BLOCK_BRANCH_ALWAYSThis function defines a trigger model block that always goes to a specific block.Type TSP-Link accessible Affected by Where saved Default valueFunction Yes Restore configurationInstrument resetPower cycleConfiguration script Not applicableUsagetrigger.model.setblock(blockNumber, trigger.BLOCK_BRANCH_ALWAYS, branchToBlock)blockNumber The sequence of the block in the trigger modelbranchToBlock The block number to execute when the trigger model reaches the Always block