Section 8: TSP command reference Model 2461 Interactive SourceMeter® Instrument Reference Manual8-320 2461-901-01 A/November 2015tsplink.line[N].reset()This function resets some of the TSP-Link trigger attributes to their factory defaults.Type TSP-Link accessible Affected by Where saved Default valueFunction YesUsagetsplink.line[N].reset()N The trigger line (1 to 3)DetailsThe tsplink.line[N].reset() function resets the following attributes to their default values:• trigger.tsplinkin[N].edge• trigger.tsplinkout[N].logic• tsplink.line[N].mode• trigger.tsplinkout[N].stimulus• trigger.tsplinkout[N].pulsewidthThis also clears trigger.tsplinkin[N].overrun.Exampletsplink.line[3].reset() Resets TSP-Link trigger line 3attributes to default values.Also seetrigger.tsplinkin[N].edge (on page 8-311)trigger.tsplinkin[N].overrun (on page 8-311)trigger.tsplinkout[N].logic (on page 8-313)trigger.tsplinkout[N].pulsewidth (on page 8-314)trigger.tsplinkout[N].stimulus (on page 8-315)tsplink.line[N].mode (on page 8-319)tsplink.line[N].stateThis attribute reads or writes the digital state of a TSP-Link synchronization line.Type TSP-Link accessible Affected by Where saved Default valueAttribute (RW) Yes Not applicable Not applicable tsplink.STATE_HIGHUsagelineState = tsplink.line[N].statetsplink.line[N].state = lineStatelineState The state of the synchronization line:• Low: tsplink.STATE_LOW• High: tsplink.STATE_HIGHN The trigger line (1 to 3)