Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI command reference2461-901-01 A/November 2015 6-211:TRIGger:DIGital:IN:CLEarThis command clears the trigger event on a digital input line.Type Affected by Where saved Default valueCommand only Not applicable Not applicable Not applicableUsage:TRIGger:DIGital:IN:CLEar Digital I/O trigger line (1 to 6)DetailsThe event detector of a trigger enters the detected state when an event is detected. For the specifiedtrigger line, this command clears the event detector, discards the history, and clears the overrunstatus(sets the overrun status to 0).For this block to work as expected, make sure you configure the trigger type and line state of thedigital line for use with the trigger model (use the digital line mode command).Example:TRIG:DIG2:IN:CLE Clears the trigger event detector on I/O line 2.Also see:DIGital:LINE:MODE (on page 6-35)Digital I/O port configuration (on page 3-87):TRIGger:DIGital:IN:OVERrun? (on page 6-212):TRIGger:DIGital:IN:EDGEThis command sets the edge used by the trigger event detector on the given trigger line.Type Affected by Where saved Default valueCommand and query Recall settingsInstrument resetPower cycleSave settings FALLUsage:TRIGger:DIGital:IN:EDGE :TRIGger:DIGital:IN:EDGE? Digital I/O trigger line (1 to 6) The trigger edge value:• Detect falling-edge triggers as inputs: FALLing• Detect rising-edge triggers as inputs: RISing• Detect either falling or rising-edge triggers as inputs: EITHerSee Details for descriptions of values