Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI command reference2461-901-01 A/November 2015 6-231:TRIGger:LOAD "SimpleLoop"This command loads a predefined trigger model configuration.Type Affected by Where saved Default valueCommand only Not applicable Not applicable Not applicableUsage:TRIGger:LOAD "SimpleLoop", :TRIGger:LOAD "SimpleLoop", , :TRIGger:LOAD "SimpleLoop", , , "":TRIGger:LOAD "SimpleLoop", , , "", The number of measurements the instrument will make The delay time before each measurement (167 ns to 10 ks); default is 0 forno delay A string that indicates the reading buffer; the default buffers (defbuffer1or defbuffer2) or the name of a user-defined buffer; if no buffer isspecified, defbuffer1 is used Define a measure or digitize block for the trigger model; options are:• ACTive: Add a measure or digitize block to the trigger model basedon the active function; if no option defined, ACTive is used• MEASure: Adds a measure block to the trigger model• DIGitize: Adds a digitize block to the trigger modelDetailsThis command sets up a loop that sets a delay, makes a measurement, and then repeats the loop thenumber of times you define in the count parameter.After selecting a trigger model template, you can view the trigger model blocks in a graphical formatby pressing the front-panel MENU key and under Trigger, selecting Configure. You can also add ordelete blocks and change trigger model settings from this screen.You can use the TRIGger:BLOCk:LIST? command to view the trigger model blocks in a list format.This command replaces the TRIGger:LOAD:LOOP:SIMPle command, which is deprecated.