Section 8: TSP command reference Model 2461 Interactive SourceMeter® Instrument Reference Manual8-176 2461-901-01 A/November 2015Examplesmu.source.func = smu.FUNC_DC_CURRENTsmu.source.autorange = smu.ONSet the source function to current.Set the instrument to select thesource range automatically.Also seesmu.source.range (on page 8-207)smu.source.autodelayThis attribute enables or disables the automatic delay that occurs when the source is turned on.Type TSP-Link accessible Affected by Where saved Default valueAttribute (RW) Yes Restore configurationInstrument resetPower cycleSource configuration listFunction changeConfiguration scriptSource configuration listsmu.ONUsagestate = smu.source.autodelaysmu.source.autodelay = statestate Disable the source auto delay: smu.OFFEnable the source auto delay: smu.ONDetailsWhen autodelay is turned on, the actual delay that is set depends on the range.When source autodelay is on, if you set a source delay, the autodelay is turned off.Examplesmu.source.autodelay = smu.OFF Turn off auto delay when current is beingsourced.Also seesmu.source.delay (on page 8-183)