Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI command reference2461-901-01 A/November 2015 6-55[:SENSe[1]]::AVERage[:STATe]This command enables or disables the averaging filter for measurements of the selected function.Type Affected by Where saved Default valueCommand and query Recall settingsInstrument resetPower cycleMeasure configuration listSave settingsMeasure configuration listOFF (0)Usage[:SENSe[1]]::AVERage[:STATe] [:SENSe[1]]::AVERage[:STATe]? The measure function:• Current: CURRent[:DC]• Resistance: RESistance• Voltage: VOLTage[:DC] The filter status; set to one of the following values:• Disable the averaging filter: OFF or 0• Enable the averaging filter: ON or 1DetailsThis command enables or disables the averaging filter. When this is enabled, the reading returned bythe instrument is an averaged value, taken from multiple measurements. The settings of the filtercount and filter type for the selected measure function determines how the reading is averaged.If you send this command without the parameter, it sets the state of the averaging filterfor all functions.Example 1CURR:AVER:COUNT 10CURR:AVER:TCON MOVCURR:AVER ONSet the averaging filter type to movingaverage, with a filter count of 10.Enable the averaging filter.Example 2RES:AVER:COUNT 10RES:AVER:TCON MOVRES:AVER ONSet the averaging filter type to movingaverage, with a filter count of 10.Enable the averaging filter.Example 3VOLT:AVER:COUNT 10VOLT:AVER:TCON MOVVOLT:AVER ONSet the averaging filter type to movingaverage, with a filter count of 10.Enable the averaging filter.Also seeFiltering measurement data (on page 4-23)[:SENSe[1]]::AVERage:COUNt (on page 6-53)[:SENSe[1]]::AVERage:TCONtrol (on page 6-56)