Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI command reference2461-901-01 A/November 2015 6-73If the NPLC setting is less than 0.2 PLC, sending autozero once can result in delay of more than asecond.This command only applies to measure functions; it does not apply to digitize functions.ExampleFUNC "VOLT"AZER:ONCEDo a one-time refresh of the reference and zeromeasurements for the voltage function.Also seeAutomatic reference measurements (on page 2-129)[:SENSe[1]]::AZERo[:STATe] (on page 6-57)[:SENSe[1]]:CONFiguration:LIST:CATalog?This command returns the name of one measure configuration list that is stored on the instrument.Type Affected by Where saved Default valueQuery only Not applicable Not applicable Not applicableUsage[:SENSe[1]]:CONFiguration:LIST:CATalog?DetailsYou can use this command to retrieve the names of measure configuration lists that are stored in theinstrument.This command returns one name each time you send it. This command returns an empty string whenthere are no more names to return. If the command returns an empty string the first time you send it,no measure configuration lists have been created for the instrument.ExampleCONF:LIST:CAT? Send this command to retrieve the name of one measureconfiguration list. To get all stored lists, send it again until itreturns an empty string.Also seeConfiguration lists (on page 3-30)[:SENSe[1]]:CONFiguration:LIST:CREate (on page 6-74)