Section 2: General operation Model 2461 Interactive SourceMeter® Instrument Reference Manual2-142 2461-901-01 A/November 2015Reset the instrumentYou can reset many of the instrument settings to their default values. For detail on what gets reset,see Reset default values (on page 4-25). Default values are also listed in the command descriptions.If you are connected to a TSP-Link system, resetting the instrument resets all TSP-Link enabledinstruments on the TSP-Link system.Using the front panel:1. Press MENU.2. Under System, select Info/Manage.3. Select System Reset.4. The commands are reset and a confirmation message is displayed.Using SCPI commands:Send the command:*RSTUsing TSP commands:Send the command:reset()If the instrument is connected to a TSP-Link system and you are using TSP commands, you canreset only the local instrument by sending localnode.reset() instead of reset().Using the event logThe event log records events, which can be errors, warnings, and information reported by theinstrument. Through the Event Log menu, you can view these events. You can also specify whichevents are shown in the event log, which ones are logged, and which ones generate popupmessages.