Section 6: SCPI command reference Model 2461 Interactive SourceMeter® Instrument Reference Manual6-110 2461-901-01 A/November 2015:SOURce[1]:PULSe:SWEep::LINearThis command sets up a linear pulse sweep for a fixed number of pulse points.Type Affected by Where saved Default valueCommand only Recall settingsInstrument resetPower cycleSave settings Not applicableUsage:SOURce[1]:PULSe:SWEep::LINear , , , ,:SOURce[1]:PULSe:SWEep::LINear , , , ,, :SOURce[1]:PULSe:SWEep::LINear , , , ,, , "":SOURce[1]:PULSe:SWEep::LINear , , , ,, , "", :SOURce[1]:PULSe:SWEep::LINear , , , ,, , "", , :SOURce[1]:PULSe:SWEep::LINear , , , ,, , "", , , :SOURce[1]:PULSe:SWEep::LINear , , , ,, , "", , , ,:SOURce[1]:PULSe:SWEep::LINear , , , ,, , "", , , ,, :SOURce[1]:PULSe:SWEep::LINear , , , ,, , "", , , ,, , :SOURce[1]:PULSe:SWEep::LINear , , , ,, , "", , , ,, , , The source function:• Voltage pulse sweep: VOLTage• Current pulse sweep: CURRent Output level the instrument sources before the first pulse and returns to betweenpulses (delay and off time):• Current: -7.35 A to 7.35 A• Voltage: -105 V to 105 V The voltage or current source level at which the pulse sweep starts:• Current: -10.5 A to 10.5 A• Voltage: -105 V to 105 V The voltage or current source level at which the pulse sweep stops:• Current: -10.5 A to 10.5 A• Voltage: -105 V to 105 V The number of pulse-measure points between the start and stop values of the pulsesweep (2 to 1e6) The time at the amplitude level for each pulse:• Extended operating area: 10 A at 100 V, 5 % duty cycle, maximum pulsewidth: 1 ms, minimum pulse width 150 μs (load dependent)• Normal operating area (DC): 99.99 % duty cycle, maximum pulse width10,000 s, minimum pulse width 150 μs