Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI command reference2461-901-01 A/November 2015 6-237Example 2*RSTTRIG:TIM4:DEL 0.5TRIG:TIM4:STAR:STIM NOT8TRIG:TIM4:STAR:GEN OFFTRIG:TIM4:COUN 20TRIG:TIM4:STAT ONTRIG:LOAD "Empty"TRIG:BLOC:BUFF:CLEAR 1, "defbuffer1"TRIG:BLOC:NOT 2, 8TRIG:BLOC:WAIT 3, TIM4TRIG:BLOC:MEAS 4, "defbuffer1"TRIG:BLOC:BRAN:COUN 5, 20, 3INIT*WAITRAC:ACT? "defbuffer1"Set trigger timer 4 to have a 0.5 s delay.Set the stimulus for trigger timer 4 to be the notify 8 event.Set the trigger timer 4 stimulus to off.Set the timer event to occur when the timer delay elapses.Set the trigger timer 4 count to 20.Enable trigger timer 4.Clear the trigger model.Set trigger model block 1 to clear the buffer.Set trigger model block 2 to generate the notify 8 event.Set trigger model block 3 to wait for trigger timer 4 to occur.Set trigger model block 4 to make a measurement and store it in default buffer 1.Set trigger model block 5 to repeat the trigger model 20 times, starting at block 3.Start the trigger model.Output the number of entries in default buffer 1.Output:20Also see:TRIGger:TIMer:CLEar (on page 6-235):TRIGger:TIMer:DELay (on page 6-238)