Section6: SCPI command referenceModel 2461 Interactive SourceMeter® Instrument Reference Manual6-2402461-901-01 A/November 2015:TRIGger:TIMer:STARt:OVERrun?This command indicates if an event was ignored because of the event detector state.TypeAffected byWhere savedDefault valueQuery onlyNot applicableNot applicableNot applicableUsage:TRIGger:TIMer:STARt:OVERrun?Trigger timer number (1 to 4)DetailsThis command indicates if an event was ignored because the event detector was already in thedetected state when the event occurred.This is an indication of the state of the event detector built into the timer itself. It does not indicate ifan overrun occurred in any other part of the trigger model or in any other construct that is monitoringthe delay completion event. It also is not an indication of a delay overrun.This returns0if there is no overrun or1if there is an overrun.ExampleTRIG:TIM1:STAR:OVER?Checks the overrun status on trigger timer 1.Also seeNone:TRIGger:TIMer:STARt:SECondsThis command configures an alarm or a time in the future when the timer will start.TypeAffected byWhere savedDefault valueCommand and queryRecall settingsInstrument resetPower cycleSave settings0Usage:TRIGger:TIMer:STARt:SEConds :TRIGger:TIMer:STARt:SEConds?Trigger timer number (1 to 4)The time: 0 to 2,147,483,647 s