Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 8: TSP command reference2461-901-01 A/November 2015 8-175smu.reset()This function turns off the output and resets the commands that begin with smu. to their default settings.Type TSP-Link accessible Affected by Where saved Default valueFunction YesUsagesmu.reset()DetailsThis function turns off the output and resets the commands that begin with smu. to their defaultsettings.Examplesmu.reset() Turns off the output and resets the SMUcommands to their default settings.Also seereset() (on page 8-97)smu.source.autorangeThis attribute determines if the range is selected manually or automatically for the selected source function orvoltage source.Type TSP-Link accessible Affected by Where saved Default valueAttribute (RW) Yes Restore configurationInstrument resetPower cycleSource configuration listFunction changeConfiguration scriptSource configuration listsmu.ONUsagesourceAutorange = smu.source.autorangesmu.source.autorange = sourceAutorangesourceAutorange Disable automatic source range: smu.OFFEnables automatic source range: smu.ONDetailsThis command indicates the state of the range for the selected source. When automatic source rangeis disabled, the source range is set manually.When automatic source range is enabled, the instrument selects the range that is most appropriatefor the value that is being sourced. The output level controls the range. If you read the range after theoutput level is set, the instrument returns the range that the instrument chose as appropriate for thatsource level.If the source range is set to a specific value from the front panel or a remote command, the setting forautomatic range is set to disabled.Only available for the current and voltage functions.