Section 2: General operation Model 2461 Interactive SourceMeter® Instrument Reference Manual2-140 2461-901-01 A/November 2015Define the setup used when power is turned onYou can select a configuration to be used when power is turned on.From the front panel:1. Set the instrument to the settings that you want it to have each time the power is turned on.2. Press the MENU key to open the main menu. Under Scripts, select Create Setup. The CREATESETUP window is displayed.3. Select Create. A keyboard is displayed.4. Enter the name of the new script, and then select ENTER on the keyboard to save it.5. The instrument saves all present system settings to the script and displays a confirmationmessage. Click OK.6. Press the EXIT key to return to the main menu.7. Under Scripts, select Run. The RUN SCRIPTS window opens.8. Select the script you just created.9. Select Copy to Power Up.10. Click OK on the confirmation message.Using a SCPI command:Send the command::SYSTem:POSetup Where is:• RST: Use the *RST defaults.• SAV0: Use the setup stored at memory location 0• SAV1: Use the setup stored at memory location 1• SAV2: Use the setup stored at memory location 2• SAV3: Use the setup stored at memory location 3• SAV4: Use the setup stored at memory location 4Using a TSP command:Save the script that you want to use as the power-on default to be autoexec. For example, to savethe commands that are presently in the instrument to be the power-on defaults, send the command:createconfigscript("autoexec")If an autoexec script already exists, you must delete it by sending thescript.delete("autoexec") command. Performing a system reset does not delete theautoexec script.