Model 2461 Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI command reference2461-901-01 A/November 2015 6-33Example:CALC2:VOLT:LIM1:CLE:AUTO OFF:CALC2:VOLT:LIM1:AUD FAIL:CALC2:VOLT:LIM1:LOW 0.25:CALC2:VOLT:LIM1:UPP 2.5:CALC2:VOLT:LIMIT1:STAT ON:READ?:CALC2:VOLT:LIMIT1:FAIL?:CALC2:VOLT:LIM1:CLESet limit autoclear off.Enable the beeper for limit 1 when a voltagemeasurement exceeds the limit.Set lower limit 1 for voltage to 0.25 V.Set upper limit 1 for voltage to 2.5 V.Enable limit 1 testing for voltage.Make a reading; the limit is checked and resultsdisplay on the front panelReturn the test results; example output if the testfails on the low limit:LOWClear the test results.Also see:CALCulate2::LIMit:CLEar:AUTO (on page 6-26):CALCulate2::LIMit:CLEar[:IMMediate] (on page 6-28):CALCulate2::LIMit:FAIL? (on page 6-29):CALCulate2::LIMit:LOWer[:DATA] (on page 6-30):CALCulate2::LIMit:UPPer[:DATA] (on page 6-33):CALCulate2::LIMit:UPPer[:DATA]This command specifies the upper limit for a limit test.Type Affected by Where saved Default valueCommand and query Recall settingsInstrument resetPower cycleMeasure configuration listSave settingsMeasure configuration list1Usage:CALCulate2::LIMit:UPPer[:DATA] :CALCulate2::LIMit:UPPer[:DATA] DEFault:CALCulate2::LIMit:UPPer[:DATA] MINimum:CALCulate2::LIMit:UPPer[:DATA] MAXimum:CALCulate2::LIMit:UPPer[:DATA]?:CALCulate2::LIMit:UPPer[:DATA]? DEFault:CALCulate2::LIMit:UPPer[:DATA]? MINimum:CALCulate2::LIMit:UPPer[:DATA]? MAXimum The measure function:• Current: CURRent[:DC]• Resistance: RESistance• Voltage: VOLTage[:DC]The digitize function:• Digitize current: DIGitize:CURRent• Digitize voltage: DIGitize:VOLTage Limit number: 1 or 2 The value of the upper limit (−9.99999e+11 to +9.99999e+11)