Appendix A: Maintenance Model 2461 Interactive SourceMeter® Instrument Reference ManualA-6 2461-901-01 A/November 2015Using TSBDo not turn off power or remove the USB flash drive until the upgrade process is complete.You can use Test Script Builder (TSB) to upgrade the firmware of your instrument.To upgrade the firmware using Test Script Builder:1. Disconnect the input and output terminals that are attached to the instrument.2. Start Test Script Builder.3. On the Instrument Console toolbar, click the Open Instrument icon.Figure 183: TSB Instrument Console toolbar4. Select your communication interface from the Select Instrument dialog box. See the Connectingan instrument in TSB (on page 7-33) for details on opening communications.5. On the Instrument Console toolbar, click the View Menu icon. Select Instrument, then selectFlash.6. From the Select a Firmware Image File dialog box, use the browser to select the file name of thenew firmware or enter the path and file name.7. If you are upgrading the firmware, replace the existing firmware with a newer version of firmware.8. If you are downgrading the firmware, replace the existing firmware with an older version offirmware or repair the same version.9. Click OK. A Progress Information bar is displayed on the instrument during the update. Inaddition, the LEDs in the upper right of the front panel blink while the upgrade is in process.10. Wait until the instrument indicates that the firmware upgrade is complete (TSB may indicate thatthe upgrade is complete before it is finalized on the instrument).11. Reboot the instrument.