Section 8: TSP command reference Model 2461 Interactive SourceMeter® Instrument Reference Manual8-218 2461-901-01 A/November 2015smu.source.userdelay[N]This attribute sets a user-defined delay that you can use in the trigger model.Type TSP-Link accessible Affected by Where saved Default valueAttribute (RW) Yes Restore configurationInstrument resetPower cycleSource configuration listFunction changeConfiguration scriptSource configuration list0UsagedelayTime = smu.source.userdelay[N]smu.source.userdelay[N] = delayTimedelayTime The delay in seconds (0 to 10 ks)N The number that identifies this user delay (1 to 5)DetailsTo use this command in a trigger model, assign the delay to the dynamic delay block.The delay is specific to the selected function.Examplesmu.source.userdelay[1] = 5trigger.model.setblock(1, trigger.BLOCK_SOURCE_OUTPUT, smu.ON)trigger.model.setblock(2, trigger.BLOCK_DELAY_DYNAMIC, trigger.USER_DELAY_S1)trigger.model.setblock(3, trigger.BLOCK_MEASURE)trigger.model.setblock(4, trigger.BLOCK_SOURCE_OUTPUT, smu.OFF)trigger.model.setblock(5, trigger.BLOCK_BRANCH_COUNTER, 10, 1)trigger.model.initiate()Set user delay for source 1 to 5 s.Set trigger block 1 to turn the source output on.Set trigger block 2 to a dynamic delay that calls source user delay 1.Set trigger block 3 to make a measurement.Set trigger block 4 to turn the source output off.Set trigger block 5 to branch to block 1 ten times.Start the trigger model.Also seetrigger.model.setblock() — trigger.BLOCK_DELAY_DYNAMIC (on page 8-288)