Section 8: TSP command reference Model 2461 Interactive SourceMeter® Instrument Reference Manual8-94 2461-901-01 A/November 2015DetailsIf startIndex is set to less than 1 or if endIndex is more than the size of the index, 9.910000e+37is returned for each value outside the allowed index and an event is generated.If overlapped commands use the specified reading buffers and the commands are not complete (atleast to the specified index), this function outputs data as it becomes available.When there are outstanding overlapped commands to acquire data, n refers to the index that the lastentry in the table has after all the readings have completed.If you pass a reading buffer instead of a reading buffer subtable, the default subtable for that readingbuffer is used.This command generates a single response message that contains all data.The format.data attribute controls the format of the response message.You can use the bufferVar attributes that are listed in the following table with the print buffercommand. For example, if testData is the buffer, you can use testData.dates attribute to printthe date of each reading in the testData buffer.You can use bufferVar.n to retrieve the number of readings in the specified reading buffer.Attribute DescriptionbufferVar.readings The readings stored in a specified reading buffer. SeebufferVar.readings (on page 8-27).bufferVar.dates The dates of readings stored in the reading buffer. SeebufferVar.dates (on page 8-20).bufferVar.statuses The status values of readings in the reading buffer. SeebufferVar.statuses (on page 8-36).bufferVar.formattedreadings The stored readings formatted as they appear on thefront-panel display. See bufferVar.formattedreadings (on page8-23).bufferVar.sourceformattedvalues The source levels formatted as they appear on the front-paneldisplay when the readings in the reading buffer were acquired.See bufferVar.sourceformattedvalues (on page 8-30).bufferVar.sourcevalues The source levels that were being output when readings in thereading buffer were acquired. See bufferVar.sourcevalues (onpage 8-33).bufferVar.sourcestatuses The source status conditions of the instrument for the readingpoint. See bufferVar.sourcestatuses (on page 8-31).