573-3-2 Laser pumping: LSRControls the laser pumping ON/OFF.⿎ Setting request data / response data for readoutStartcode LSR Sub command [Laser pumping control]1-byte (Check sum) End codeSub command is “S” for the setting request data and “A” for the response data for readout.⿎ Readout request dataStartcode LSR R (Check sum) End code⿎ Data descriptionName Data length [byte] Data contents RemarksLaser pumpingcontrol 1 “0”: Laser pumping OFF“1”: Laser pumping ONTotal datalength 1 The data length is fixed.ンㄆㄇㄆㄓㄆㄏㄆ• This command is available only when the laser pumping control method is set to “communication commands” by LaserMarker NAVI smart.• From sending the laser pumping on request to completion of laser pumping, the following time will be required.• LP-GS series: about 8 to 15 seconds• LP-RC series: about 10 seconds• LP-RF series: about 7 seconds• LP-RV series: about 1 second• For LP-RC series, the shutter can be opened after turning on the laser pumping. When the laser pumping is turned tooff, the shutter is closed automatically.• Response data when reading out this command returns the laser pumping control status. If you want to check the laserpumping ON/OFF status, use the status checking (STS) command.• When you turn OFF the laser pumping during marking laser processing (laser radiation), an error occurs and thisinterrupts the marking trigger processing.⿎ Example of command settingsCondition: Laser pumping control = 1• Setting Transmission data [STX] LSR S 1 [CR]• Readout Transmission data [STX] LSR R [CR]Response data [STX] LSR A 1 [CR]ME-LP-GS-SR-COMP-3