503-2 Start-up & TerminationWARNING• It is obligated by IEC/FDA/JIS that laser products shall incorporate a key-actuated master control. Actuation of this laser marker is basically controlledby the key switch located on the front of the controller box. However, inconsidering situations when the laser marker is operating as a part of a largersystem, the laser marker turns on if the key switch is already in ON position,and power is supplied. In this case, be sure that the external system controlsthe operation of the laser marker with a key-actuated master control.3-2-1 Start-up procedure1. Setup the laser marker connecting head, controller and interfaces.Refer to “2-5 Connecting Laser Marker” (P.41).2. Turn ON the key switch of the controller.The MAIN power indicator of the controller flashes.3. The system startup completes approximately after 10 seconds.The MAIN power indicator of the controller changes from blinking light tosteady light.ワㄐㄕㄊㄆ• Do not set the key switch to OFF until system startup has been completed.• In case of turning ON the power supply after turning OFF, leave the interval five seconds or more between ON and OFF.4. To configure and control by the PC configuration software, start Laser Marker NAVI smart andestablish an online connection with the laser marker.Click the “Start” button on Windows. Go to “All Programs” (for Windows 8 or Windows 10, go to “All apps”), clickin the order of “Panasonic-ID SUNX Laser” - “Laser Marker Smart Utility” - “Laser Marker NAVI smart” to start theapplication.Refer to “3-3-3 How to establish online connection” (P.56).ヮモリワロモヴユンヮモンレリワヨモロモンヮヱヰヸユンONOFFME-LPGS-SM-8