3-10 BOP HIPWR 111315To change the background, press # from the power-up screen, Highlight Background, press!, highlight Black or White, then $ to save. Press $ exit and save for power-up or % toapply the changes (without saving for power-up) and exit.3.2.5.2 DISPLAYING PROGRAMMED SETTINGS IN REMOTE MODEWhen Remote Debug is set to OFF (default), the programmed settings for main and protectionchannels that are displayed on the LCD in local mode, are not displayed while in remote modeand the LCD reads “Display settings off.” When set to ON, the programmed settings for mainand protection channels are displayed, however the unit responds more slowly in remote mode.To change the Remote Debug setting, press # from the power-up screen (Figure 3-3), High-light Remote Settings, press !, highlight On or Off, then $ to save. Press $ exit and savefor power-up or % to apply the changes (without saving for power-up) and exit.TABLE 3-3. DISPLAY (OPERATOR CONVENIENCE) MENU FUNCTIONSSETTING CHOICES(BOLD = Factory Default) FUNCTIONNOTE: To save any of the functions listed below for power-up it is necessary to first change SAVE DISPLAY CHG at the passwordsetup (see PAR. 3.2.4.2and Figure 3-5) from DISABLE to ENABLE.GRAPHIC DISPLAY Meters75s Graph30s Graph15s Graph7.5s Graph3.8s Graph1.5s Graph0.8s Graph0.5s Graph0.3s Graph0.15s Graph0.04s Graph0.08s GraphMETERS - Output voltage and current are displayed on representations of ana-log meters.GRAPH - Displays a sample of the output for the time indicated (see Figure 3-4). The complete time-line uses 240 points. E.g., selecting 2 Second Graphmeans that 2S/240 = 50ms sampling, i.e., every 50ms a snapshot of the outputstate is used to produce a single point.DISPLAYBACKGROUNDWhiteBlackWHITE - Sets the background to white and displays black characters.BLACK - Sets the background to black and displays white characters.REMOTEDEBUGOffOnOFF - LCD does not show programmed settings for main and protection chan-nels while in remote mode of operation (local mode is unaffected).ON - Programmed settings for main and protection channels are displayed,however the unit responds more slowly.KEY PRESS BEEP OnError OnlyOffON - provides an audible beep for each key press and for each error received.ERROR ONLY - silences the beeps for each key press; errors will still causebeep.OFF - silences all beeps including key presses and errors except for power-upand unit faults.WAVEFORM STOP Output OffLast LevelCurrent LvlOutput Off - When waveform is stopped (either count complete or %pressed), output freezes at whatever level the waveform was at when % waspressed or count completed, then after a brief delay output is set to STANDBY(disabled).Last Level - When waveform is stopped (either count complete or %pressed), waveform runs through full cycle. The output stops at the last gener-ated value of the waveform (e.g., for a sinewave segment of 0° to 270°, 10Vamplitude, 0V offset, when F5 is pressed the output will end up at -10V).Current lvl - When waveform is stopped (% pressed), output freezes at what-ever level the waveform was at when % was pressed.NOTE: Control display contrast by pressing # twice to dynamically change the contrast, then press # to set when contrast isacceptable. Contrast can also be set directly by pressing # once and using T and R as needed (see PAR. 3.2.5.1).