3-34 BOP HIPWR 111315FIGURE 3-11. SAMPLE WAVEFORM2. The initial waveform settings are displayed (see Table 3-11 and Table 3-12 for parameterdetails. Use ! to modify the initial settings for the waveform. For each parameter, use $to save the setting.Name CAP CHARGMode VOLTAGEProtectionPositive 10Negative 10Count 0SegmentType Neg RampFrequency (Hz) 50Amplitude (p-p) 4Offset 8When all settings are complete, press $ to save for power-up. This step creates a repeat-ing negative ramp starting at +10V decreasing to +6V since it is a 4V p-p signal riding on an8V offset. The frequency of 50 Hz establishes the time duration of 20 ms for the segment.3. From the Waveform Settings screen, highlight segment NEG RAMP 50.0HZ 4.00 R andpress @ to insert a new segment. Use ! to modify the following settings of the new seg-ment. For each parameter, use $ to save the setting.Type LevelPeriod (Sec) 0.002Offset 0Initial/Repeat InitialWhen complete, press $ to save for power-up. This step changes the 2ms, 0V level to beexecuted only once at the beginning of the waveform.4. From the Waveform Settings screen, highlight segment LEVEL 0.002s 0.00 I and press @to insert a new segment. Use ! to modify the following settings of the new segment. Foreach parameter, use $ to save the setting.Type Level