8.6.5 MIP PolarimeterMeasPlotThis VI can be used as standalone VI to start a measurement and displaythe results in a graph. All important parameters can be set prior to themeasurement using the defined control elements. Changes in the controlelements are valid only after restarting the VI.Input parameters (con-trol elements):VISA descriptor, Baudrate, measurementmode, trigger source, sampling rate, num-ber of samples, wavelength, sweeprate,gainOutput parameters(display on screen):Results plot, error codeTriggerSource• NONE (starts immediately; software trigger)• TTL HIGH (low-to-high transition on external trigger-in startsmeasurement)• TTL LOW (high-to-low transition on external trigger-in starts mea-surement)SamplingRate• Sampling rate in kHzWavelength• Lightsource wavelength in nmSweepRate• Lightsource wavelength sweeprate in nm/sNOTE: for fixed wavelength source set to 0 nm/sGain• Amplifier gain (0-13)8.6.6 MIP PolarimeterResetThis VI is used to reset the polarimeter into its initial state. All mea-surement parameters are set to their values after device power-up.72 A1XXX/A2XXX/A3XXX User Guide