15 Reference15.1 Command Line ParametersThe following command line parameters are supported by the software:Command Line Switch Description/minimize The software will be minimized to the sys-tem tray on startup. This may be usefulif the polarizationNAVIGATORTM is onlyneeded for remote control or as automationtarget.15.2 Polarimeter Input RangesThe following table shows the permitted input power ranges for all avail-able gain settings. If the input power falls below the lower boundarymeasurement results become noisy due to quantization. If the inputpower exceeds the upper boundary, errors may appear due to limits ofthe input amplifier. You can manually choose a specific gain setting bydisabling the auto-gain function (i.e. TreeNo=3/VarNo=14 (section 10.2)AutoGainFlag = 0).Gain Max. Input Power Power Range0 7 dBm 0dBm .. 7dBm1 4 dBm -3dBm .. 4dBm2 0 dBm -7dBm .. 0dBm3 -3 dBm -10dBm .. -3dBm4 -6 dBm -13dBm .. -6dBm5 -10 dBm -17dBm .. -10dBm6 -13 dBm -20dBm .. -13dBm7 -16 dBm -23dBm .. -16dBm8 -20 dBm -27dBm .. -20dBm9 -23 dBm -30dBm .. -23dBm10 -26 dBm -33dBm .. -26dBm11 -30 dBm -37dBm .. -30dBm12 -33 dBm -40dBm .. -33dBm13 -36 dBm -43dBm .. -36dBmA1XXX/A2XXX/A3XXX User Guide 127