Keysight N9923-90001 User’s Guide67Time Domain – Option 010Time Domain (Transform) SettingsThe Window setting applies to ALL traces.How to make Window setting— Press Measure Setup 4— Then Transform— Then More— Then Transform Window— A settings table appears that allows you to navigate and select thefollowing settings. These settings are made exactly like those made on theTransform Settings table. Learn how in “Transform Settings Table” onpage 64.Window SettingsWindow settings provide the ability to choose between optimizing TDRmeasurements for resolving closely-spaced faults or for the ability to measurelow-level faults.Any of the following four methods can be used to make this same Windowsetting. Impulse Width values are calculated from the frequency span andKaiser Beta value.1. Window Min Medium Max2. Percent 0 50 1003. Kaiser Beta 0 6.50 13.004. Impulse Width 100.455 ps 168.501 231.284