Keysight N9923-90001 User’s Guide235System SettingsService DiagnosticsService Diagnostics— The Service diagnostics menu is generally used to display service-relatedinformation. To access these menu items, press System 7 then ServiceDiagnostics.System InformationThe System information table shows important information about yourFieldFox. The following is the most useful information:— Model Number – FieldFox model— Serial Number – Identification number of your FieldFox— Firmware Release – The currently installed version of software. Check tosee if you have the latest firmware at:www.keysight.com/find/fieldfoxsupport— Network Identity – The network name and current IP address of yourFieldFox.— Reflectometer Factory Cal - The Factory (CalReady) Calibration shows thedate and time that the factory cal was performed for each test port and theThru connection. The format is P1 , P2 , P1-P2. The factory calibration is updated when you send yourFieldFox to Keysight for Instrument Calibration. Learn more in “InstrumentCalibration” on page 265.Event LogThe Event Log shows the information that has appeared on the FieldFoxscreen. The most recent information is at the top of the table.Press Clear Events to erase this information.BatteryLearn more in Chapter 16, “Working with the Lithium-Ion Battery”, on page239.Internal TemperaturesLearn about the importance of the internal FieldFox temperatures in “FieldFoxHigh Temperature Protection” on page 24.Erase User DataAll data that you have written to the FieldFox can be erased with the press of abutton. You would do this if your FieldFox is located in a secure environmentand must be routinely erased of sensitive data.