Keysight N9923-90001 User’s Guide53NA (Network Analyzer) ModeNA Mode Settings— Then choose from the following:101 |201 | 401 | 601 | 801 | 1001 | 1601 | 4001 | 10001.— Using SCPI, Resolution can be set to ANY number of points between 3 and10001. See the Programming Guide athttp://na.support.keysight.com/fieldfox/help/Programming/webhelp/FFProgrammingHelp.htmSweep TimeThe fastest possible sweep time is always used as the default setting. Use theMin Swp Time setting to slow the sweep time when measuring long lengths ofcable. Learn more at the FieldFox Supplemental Online Help:http://na.support.keysight.com/fieldfox/help/SupHelp/FieldFox.htmThe actual sweep time is shown on the FieldFox screen. See the Screen Touron “Screen Tour” on page 32. To increase the sweep time, enter a value that ishigher than the actual sweep time. The in crease will not be exactly the amountthat you enter, as the actual sweep time is the composite of many factors.How to set Sweep Time— Press Sweep 3.— Then Min Swp Time.— Enter a value using the numeric keypad.— Press a multiplier key. Learn about multiplier abbreviations in “MultiplierAbbreviations” on page 34.Output PowerSet the power level out of the FieldFox to High, Low, or manually set powerlevel to a value between High and Low.Generally, the high power setting is used when measuring passive, high-lossdevices to place the signal farther from the noise floor. However, for devicesthat are sensitive to high power levels such as amplifiers, use the Low powersetting.For best measurement accuracy, use the Manual power setting at -15 dBm.After calibration, the power level can be decreased for amplifiers, or increasedfor higher dynamic range.Measurement speed specifications do NOT apply in Temperature ControlMode. Learn more in “Temperature Control Mode” on page 25.Power Level settings in this mode will NOT change Power Level settings inother modes. To help prevent damage to your DUT, use caution whenchanging modes with your DUT connected to the FieldFox test ports.