178 Keysig ht N9923-90001 User’s GuideData Analysis FeaturesAll about MarkersHow to move a marker between data and memory traces— Create a data and memory trace that will host the marker. Learn how in“How to display data and memory traces” on page 189.— Press Marker and Enter to create a marker.— Then More.— Then select the marker to move using Marker 1,2,3,4,5,6.— Thentoggle Marker Trace Auto Data Mem.—Auto - Marker resides on data trace when it is visible. Marker resideson memory trace when only IT is visible (Default setting).—Data - Marker resides on data trace, even when not visible.—Mem - Marker resides on memory trace, even when not visible.Marker FormatThe Marker Format setting, available ONLY in NA Mode, allows you to choosethe format of the marker readout that appears in the upper-right corner of thescreen and in the marker table. The marker format can be independent of thetrace format, and can be set independently for each trace. All markers on atrace have the same marker format.How to choose marker format— Press Marker— Then More— Then Marker Format— Then choose from the following:Selection Example Y-axis valueDefault Same as displayed formatR + jX Complex impedance format(3rd number is distance for Time Domain)22.8Ω -j61.4Ω287.9fFZ Magnitude Impedance Mag. 66.08ΩPhase 73.8°Real 0.2003MoreImaginary -0.6727Mag & Phase 0.705, -73.7°