66 Keysight N9923-90001 User’s GuideTime Domain – Option 010Time Domain (Transform) Settings— Then choose from:—Start Sets the transform start time.—Stop Sets the transform stop time.Distance UnitsWhen markers are present on a Time Domain trace, marker X-axis values areshown both in units of time (seconds) and distance. Select the units in whichthe marker distance data is presented.Distance Units are also used when setting Time Domain Line Loss. Learn morein. “Distance Units” on page 66.Choose from: m (meters), or Feet.Window LayoutThis setting is used to create additional traces, or select a new layout for tracesthat are already created,Choose from: x1, x2, x2H, x3H, x3, x4H.Learn more about multi-trace configurations in “Multi-Trace Configurations”on page 43.Transform WindowThere are abrupt transitions in a frequency domain measurement at the startand stop frequencies, causing overshoot and ringing in a Time Domainresponse. The Window setting reduces th e abruptness of the frequency domaintransitions. This causes you to make a trade off in the Time Domain response.Zero (0) seconds is always the calibration reference plane. Negative valuesare useful if moving the reference plane.