Series 2280 Precision Measurement DC Power Supplies Reference Manual Appendix B: Verification077085501 / February 2015 B-3DC voltage setting accuracy with remote senseModel numberDUTvoltage (V)Testcurrent (A) Min (V) Max (V)Measuredvoltage (V)0 % test voltage2280S-32-6 0 3 -0.003 0.0032280S-60-3 0 1.6 -0.006 0.00625 % test voltage2280S-32-6 8 3 7.99684 8.003162280S-60-3 15 1.6 14.9937 15.006350 % test voltage2280S-32-6 16 3 15.99668 16.003322280S-60-3 30 1.6 29.9934 30.006675 % test voltage2280S-32-6 24 3 23.99652 24.003482280S-60-3 45 1.6 44.9931 45.0069100 % test voltage2280S-32-6 32 3 31.99636 32.003642280S-60-3 60 1.6 59.9928 60.0072DC voltage setting accuracy without remote senseModel numberDUTvoltage (V)Testcurrent (A) Min (V) Max (V)Measuredvoltage (V)0 % test voltage2280S-32-6 0 3 -0.0045 0.00452280S-60-3 0 1.6 -0.0068 0.006825 % test voltage2280S-32-6 8 3 7.9939 8.00612280S-60-3 15 1.6 14.9932 15.006850 % test voltage2280S-32-6 16 3 15.9923 16.00772280S-60-3 30 1.6 29.9902 30.009875 % test voltage2280S-32-6 24 3 23.9907 24.00932280S-60-3 45 1.6 44.9872 45.0128100 % test voltage2280S-32-6 32 3 31.9891 32.01092280S-60-3 60 1.6 59.9842 60.0158