Series 2280 Precision Measurement DC Power Supplies Reference Manual Section 7: SCPI command reference077085501 / February 2015 7-33:CALibration:PROTected:COUNt?This query command returns the number of times that the instrument has completed calibration.Type Affected by Where saved Default valueQuery only Not applicable Nonvolatile memory Not applicableUsage:CALibration:PROTected:COUNt?DetailsReturns the number of times that the instrument has completed calibration. Only after saving thecalibration constants, the calibration count will be incremented.Example:CAL:PROT:COUN? Return the number of times the calibration wascompleted.Also seeNone:CALibration:PROTected:DATA?This command queries the saved calibration constants.Type Affected by Where saved Default valueQuery only Not applicable Nonvolatile memory Not applicableUsage:CALibration:PROTected:DATA?DetailsFor details, refer to Calibration constants (on page C-22).Example:CAL:PROT:DATA? Return the calibration constants in order:+4.989000E-02,+4.997200E+00,+9.091379E-01,+6.047600E-01,+1.306331E-02,+5.688220E+01,+1.230724E+00,+1.521200E-03,+4.262407E-03,+4.687520E-03,+1.500000E-01,+3.965609E-01,+3.980984E-03,+3.517675E-01,+5.200000E-03,+1.376206E-02,+1.458743E-02,+5.031000E-01,+1.332121E+00Also seeNone