Section 7: SCPI command reference Series 2280 Precision Measurement DC Power Supplies Reference Manual7-110 077085501 / February 2015:TRACe[n]:CLEar or :DATA[n]:CLEarThis command clears all readings in the buffer.Type Affected by Where saved Default valueCommand only Not applicable Not applicable Not applicableUsage:TRACe[n]:CLEar:DATA[n]:CLEar[n] Channel number: For single-channel instruments, the value is 1.Example:TRACe:CLEar Clear the data buffer.Also seeReading buffer (on page 3-7):TRACe[n]:CLEar:AUTO or :DATA[n]:CLEar:AUTOThis command enables or disables the autoclear feature for the buffer.Type Affected by Where saved Default valueCommand and query Power cycle Not applicable 1 (ON)Usage:TRACe[n]:CLEar:AUTO :TRACe[n]:CLEar:AUTO?:DATA[n]:CLEar:AUTO :DATA[n]:CLEar:AUTO?[n] Channel number: For single-channel instruments, the value is 1. Enable autoclear: 1 or ONDisable autoclear: 0 or OFFDetailsWith autoclear enabled, the buffer will automatically clear when the storage process starts. Whendisabled, readings will append to old readings in the buffer until the buffer becomes full (2,500readings) or the storage process is stopped.Example:TRACe:CLEar:AUTO ON Enable autoclear for the buffer.Also seeReading buffer (on page 3-7)