Section 7: SCPI command reference Series 2280 Precision Measurement DC Power Supplies Reference Manual7-64 077085501 / February 2015[:SOURce[n]]:DELayThis command sets the source delay.Type Affected by Where saved Default valueCommand and query Recall settingsInstrument resetPower cycleSave settings 0.001Usage[:SOURce[n]]:DELay [:SOURce[n]]:DELay?[n] Channel number: For single-channel instruments, the value is 1. The delay in seconds: 0 to 99 s. The resolution is 0.00001 s.DetailsThis command sets a delay for the source. After the programmed source is turned on, this delayallows the source level to settle before a measurement is taken. For details, refer to Source delay (onpage 3-6).This delay is the same for both the current and voltage source. Delay time must be either zero or apositive value. If you attempt to set it to a negative value, an error -222, "Data out of range" isgenerated.Example:DELay 0.05 Set the delay for the output source to 50 ms.Also seeNone[:SOURce[n]]:DELay:STATeThis command enables or disables the delay that occurs when the source is turned on.Type Affected by Where saved Default valueCommand and query Recall settingsInstrument resetPower cycleSave settings 1 (ON)Usage[:SOURce[n]]:DELay:STATe [:SOURce[n]]:DELay:STATe? Channel number: For single-channel instruments, the value is 1. Disable the source delay: OFF or 0Enable the source delay: ON or 1