Section 7: SCPI command reference Series 2280 Precision Measurement DC Power Supplies Reference Manual7-90 077085501 / February 2015DetailsThis command sets or reads the contents of the enable register of the Operation InstrumentSummary Event Register (OISER).When one of these bits is set and the corresponding bit in the OISER is set, bit 1 in the OperationInstrument Event Register is set.When sending binary values, preface with #b. When sending hexadecimal values, preface with #h. No preface is needed when sending decimal values.Example:STAT:OPER:INST:ISUM:ENAB #b0000000100010000 Sets the filter and measure bits of theOperation Instrument Summary EnableRegister using a binary value.You could also send the decimal value272 or the hexadecimal value #h110Also seeOperation Instrument summary Event Register (on page D-11):STATus:OPERation:INSTrument:ISUMmary:CONDition?This command reads the Operation Instrument Summary Condition Register of the status model.Type Affected by Where saved Default valueQuery only Not applicable Not applicable Not applicableUsage:STATus:OPERation:INSTrument:ISUMmary:CONDition?DetailsThis command reads the contents of the Operation Instrument Summary Condition Register.For details about interpreting the value of a register, see Operation Instrument Summary EventRegister (on page D-11).Example:STAT:OPER:INST:ISUM:COND? Returns the contents of the OperationInstrument Summary Condition Register ofthe status model.Also seeOperation Instrument Summary Event Register (on page D-11)