Section7:SCPI command referenceSeries 2280Precision Measurement DC Power SuppliesReference Manual7-18077085501 / February 2015:ARM[:SEQuence[n]]:COUNtThis command sets the number of times to repeat the Arm event.TypeAffected byWhere savedDefault valueCommand and queryRecall settingsInstrument resetPower cycleSave settings1Usage:ARM[:SEQuence[n]]:COUNt:ARM[:SEQuence[n]]:COUNt?:ARM[:SEQuence[n]]:COUNt?DEFault:ARM[:SEQuence[n]]:COUNt?MINimum:ARM[:SEQuence[n]]:COUNt?MAXimum[n]Channel number: For single-channel instruments, the value is1.The number of times to repeat the arm event: 1 to 2500.ExampleARM:COUN4Set the arm count to four.Also see:TRIGger[:SEQuence[n]]:COUNt(on page7-123)Trigger model(on page3-21):ARM[:SEQuence[n]]:SOURceThis command selects the control source for the arm layer.TypeAffected byWhere savedDefault valueCommand and queryRecall settingsInstrument resetPower cycleSave settingsIMMUsage:ARM[:SEQuence[n]]:SOURce:ARM[:SEQuence[n]]:SOURce?[n]Channel number: For single-channel instruments, the value is1.The trigger source for the arm layer:Immediate triggering; no delay or gating in arm detection layer:IMMediate.External triggering:EXTernal.Manual triggering; the trigger signal is generated by pressing theTRIGGERkey:MANual.Bus triggering by the command*TRG:BUS.