Series 2280 Precision Measurement DC Power Supplies Reference Manual Section 7: SCPI command reference077085501 / February 2015 7-97DetailsThis command enables or disables the beeper for error messages. When enabled, a beep signalsthat an error has happened.Example:SYSTem:BEEPer:ERRor ON Enable the error beeper.Also see:SYSTem:KCLick (on page 7-103):SYSTem:COMMunication:ABORtThis command causes the remote control to transfer to the present interface.Type Affected by Where saved Default valueCommand only Not applicable Not applicable Not applicableUsageSYSTem:COMMunication:ABORtDetailsThe instrument supports the following interfaces: USB LAN GPIB LXIOnly one interface may be connected at a time. When one interface is active, all the others areblocked.When this command is sent by one of the interfaces, the present active interface is blocked, and theinterface that sends SYST:COMM:ABOR becomes active.Before you use this command to get control of the instrument, make sure this interface is unlocked. Ifthe interface is locked, sending SYST:COMM:ABOR causes error 511, "Communication not permitted:This interface is locked, send SYSTem:PASSword:UNLock with a password to unlock it."Example:SYSTem:COMMunication:ABORt For example, if the present active interface is USB,send this command through GPIB. After that, GPIBis activated and USB is blocked.Also see:SYSTem:PASSword:MODE (on page 7-105):SYSTem:PASSword:UNLock (on page 7-106)