Series 2280 Precision Measurement DC Power Supplies Reference Manual Section 7: SCPI command reference077085501 / February 2015 7-49Also see:OUTPut:DELay:FALLing (on page 7-47):OUTPut[:STATe] (on page 7-51):OUTPut:DELay:STATeThis command enables and disables the delay time that the instrument waits before enabling and disabling thespecified output.Type Affected by Where saved Default valueCommand and query Recall settingsInstrument resetPower cycleSave settings 0 (OFF)Usage:OUTPut:DELay:STATe :OUTPut:DELay:STATe? Disable output delay: 0 or OFFEnable output delay: 1 or ONDetailsWhen output delay is set to ON, the instrument waits the specified delay time before turning on andoff the output. For details, refer to Output delay (on page 3-5).Separate delays can be programmed for the off-to-on and the on-to-off transition using the followingcommands: :OUTPut:DELay:RISing :OUTPut:DELay:FALLingExample:OUTP:DEL:STAT ON Enable the output delayAlso see:OUTPut:DELay:FALLing (on page 7-47):OUTPut:DELay:RISing (on page 7-48):OUTPut:PROTection:CLEarThis command clears the overvoltage, overcurrent, overtemperature, or reverse fault status.Type Affected by Where saved Default valueCommand only Not applicable Not applicable Not applicableUsage:OUTPut:PROTection:CLEar:OUTPut:PROTection:CLEar The channel number: Channel 1: CH1 Channel 2: CH2The default value is CH1