Series 2280 Precision Measurement DC Power Supplies Reference Manual Section 7: SCPI command reference077085501 / February 2015 7-55:SENSe[n]::AVERage:WINDowThis command specifies a filter window to control filter threshold.Type Affected by Where saved Default valueCommand and query Recall settingsInstrument resetPower cycleSave settings 1Usage:SENSe[n]::AVERage:WINDow :SENSe[n]::AVERage:WINDow?:SENSe[n]::AVERage:WINDow? DEFault:SENSe[n]::AVERage:WINDow? MAXimum:SENSe[n]::AVERage:WINDow? MINimum Channel number: For single-channel instruments, the value is 1. The measurement function to which this setting applies: Concurrent: CONCurrent[:DC] Current: CURRent[:DC] Voltage: VOLTage[:DC] The filter window: 0.01 to 100. This value is expressed in terms of a percentage.There are five acceptable window percentages: 0.01, 0.1, 1, 10, and 100.DetailsThe <value> parameter can be set to any value from 0.01 to 100. The instrument rounds up thevalue to the nearest window value. For example, the filter window is set to 1 if you send the followingcommand:SENS:CONC:AVER:WIND 0.5ExampleSENS:CONC:AVER:WIND 0.1 Set the average filter window as 0.1%.Also see:SENSe[n]::AVERage:COUNt (on page 7-52):SENSe[n]::AVERage[:STATe] (on page 7-53):SENSe[n]::AVERage:TCONtrol (on page 7-54)