Section 9: Keithley External Control Interface (KXCI) Model 4200A-SCS Parameter Analyzer Reference Manual9-4 4200A-901-01 Rev. C / February 2017Setting the remote control mode (GPIB or ethernet)Set the remote control mode on the 4200A-SCS through KCon. By default, the 4200A-SCS is set upfor GPIB remote control.To set up the remote control mode on the 4200A-SCS:1. Open KCon.2. Select KXCI Settings. The KXCI Settings dialog box is displayed, as shown in the followingfigure.3. Select the Command Set. You can select the Keithley 4200A (Normal) (the default) or Keysight4145 Emulation. In many cases, test programs developed for use with a 4145B ParameterAnalyzer run without modification when they are used with a 4200A-SCS running KXCI.Figure 461: KXCI Settings dialog boxTo set up GPIB communications:1. Select GPIB.2. Set the GPIB Address.3. Select a Reading Delimiter. The delimiter determines if the delimiter for output data is a stringterminator or a comma.4. If you selected String Terminator, select the String Terminator. The string terminator determinesif an EOI (End or Identify) signal is sent after data transfer.To set up ethernet communications:1. Select Ethernet.2. Set the Port Number (default is 1225).