Alarm, error, and system messages7.5 Wizard-based error messagesOperating with the Local User InterfaceOperating Manual, 06/2017, A5E31930478-05 2737.5 Wizard-based error messagesOverviewThe following tables include descriptions of errors which may occur when executing wizard-based processes. The wizard-based procedures especially include calibrations and AutoCalprocedures.NoteRepresentation of wizard-based error messages in the logbookErrors which prevent starting the wizard do not generate an entry in the logbook.Errors which occur during execution of a wizard generate an entry in the logbook. However,only the corresponding error number is entered in this case, not the message text.Error messagesTable 7- 8 Wizard-based error messages (F1001, ... , F1031)Error ID Message text (cause/remedy)F1001 to F1004 Internal error occurred.F1006 "F1006: The redetermined zero point and spanpoint deviate too much from the last zero pointand span point determined. The maximum per-mitted deviation is set in the menu Calibrationtolerance. Possible causes:1. An incorrect gas,2. Faulty switching of the solenoid valves,3. Different reference gas without new standardi-zation or4. A faulty analyzer module."F1007 "F1007: The measured or calculated parameter isoutside of the permitted min/max limits. Checkyour entries and setpoints."F1008 "F1008: The gas flow is too low.Check to see if the gas bottle is open or if thecorrect valve is connected and turned on."F1009 "F1009: Incorrect calibration gas detected."F1010 "F1010: There are active errors. Correct thepending errors and restart the wizard."F1012 "F1012: Internal error occurred."F1013 "F1013: A component could not be found. Thewizard cannot be started."F1014 "F1014: The calibration type does not match theselected components. Check the selection ofcomponents or change the calibration type."F1015, ... , F1017 "F101n: Internal error occurred. The wizard can-not be started."