Functions6.14 [2.20] Settings > ServiceOperating with the Local User InterfaceOperating Manual, 06/2017, A5E31930478-05 201The current measured value is output in the unit "ppm" ("Measured value in ppm"display field) and in the user-defined unit ("Meas. value in act. unit").– Unit not "ppm" or "User-defined"When you set this unit, the conversion factor for the unit "ppm" is displayed.The current offset (actual unit) and the measured value in the unit "ppm" and in thecurrently set unit are also displayed.6.14.6 [2.20.11] Physical measuring rangeService menu. You only have read access.This menu contains all information on the physical measuring range of the component. Inparticular, it includes information on the high and low range limits as well as the smallestspan. The displays are the same as those for OXYMAT 7 and ULTRAMAT 7. The displayrange is expanded for CALOMAT 7:Table 6- 50 Display scope menu [2.20.11]Displayed pa-rametersOXYMAT7ULTRAMAT7CALOMAT7NotesHigh measuringrange limitYes Yes Yes Maximum permissible measuredvalue (full-scale)Permitted highlimit violationNo No Yes Maximum permitted violation ofthe upper measuring range limit.This value is specified as a per-centage of the maximum measur-ing range span 1) .Smallest span,highNo No Yes Smallest permitted measuredvalue span for the upper measur-ing range limit in which the accu-racy is specified.Low range limit Yes Yes Yes Minimum permissible measuredvalue (characteristic initial value)Permitted low limitviolationNo No Yes maximum permitted violation ofthe lower measuring range limit.This value is specified as a per-centage of the maximum measur-ing range span 1) .Smallest span, low No No Yes Smallest permitted measuredvalue span for the lower measur-ing range limit in which the accu-racy is specified.Smallest span Yes Yes No Minimum permissible span inwhich the accuracy is specified1) Maximum measuring range span = upper measuring range limit - lower measuring range limit