Functions6.14 [2.20] Settings > ServiceOperating with the Local User Interface200 Operating Manual, 06/2017, A5E31930478-056.14.5 [2.20.10] Name/unitProcedure1. Main menu > "2 Settings" > "20. Service" > "10. Name/unit"This menu is used for setting the component name and to select the appropriate unit.NoteComponent names with several meanings/assigned several timesIf you assign names several times, the components cannot be distinguished in the mainview and during navigation.• Assign only unique component names, for example, "O2(A)" and "O2(B)" for a devicewith two OXYMAT 7 modules.• Make sure that you do not assign multiple names.2. Enter the name of the measured component: "Component name" parameter field– Authorize access with Expert PIN, if necessary.– The component name can have a maximum of 31 characters. But only sevencharacters will be visible on the display.– Enter the name with no more than seven characters.3. Specify the measurement unit of the component: "Unit" parameter field.NoteReference temperature during conversion of unitsWhen converting the measured values to volume-based units, e.g., vpm, mg/m³, thereference temperature is 0 °C.Select one of the units included in the selection list.– Selection "ppm"When you set this unit, the conversion factor for the conversion into the unit "ppm" isdisplayed.The current offset and the measured value in the unit "ppm" are also displayed.– Selection "User-defined"Additional parameter fields are displayed with this selection:Parameter field Action"User-defined unit" Assign name of unit, up to seven characters."Factor" Enter conversion factor for conversion to "ppm"."Offset" Enter offset for conversion of user-defined unit to "ppm".