Functions6.14 [2.20] Settings > ServiceOperating with the Local User InterfaceOperating Manual, 06/2017, A5E31930478-05 195Executing normalizationNoteMeasured value errorsIf you change the setpoints but do not normalize the module, the changed values are savedand the measured value is calculated incorrectly.After you have changed one or both setpoint settings, always run a normalization!1. Main menu > "2. Settings" > "20. Service > "03. Factory calibrations" > "03.Normalization".The following current values are displayed in the corresponding display fields:– Factor for determining the intensity: "k norm intensity" display field– Snorm measured value at minimum signal: "S norm offset" display field– Factor for determining C norm : "k norm span" display field– The current phase: "Phase" display field2. Enter setpoints:– Authorize access with Expert PIN, if necessary.– Enter setpoints: Use the parameter fields "Setpoint zero value" and "Setpoint endvalue".3. Start normalization calibration wizard: "Start normalization" parameter field. After thewizard has run, the following newly determined values are displayed:– "k norm intensity"– "S norm offset"– "k norm span"– "Phase"6.14.2.5 [2.20.03] Factory calibrations CALOMAT 7Executing normalization calibrationNoteRight to change setpointsThe analyzer module works with pure gases and is matched to the local application in thefactory. Make sure that the settings for the setpoints are only changed by suitably trainedpersonnel.